บริษัท จรัญเอสโซซิเอทส์ จำกัด

Flame Photometer Model: BWB-NUCLEAR

การวัดค่าลิเธียมที่ระดับความเข้มข้นต่ำภายในโรงงานไฟฟ้าพลังงานนิวเคลียร์นั้น มีความสมบูรณ์มากขึ้นด้วยเครื่องเฟลมโฟโตมิเตอร์เฉพาะกิจอย่าง The
BWB Nuclear.

ยี่ห้อ: BWB        ประเทศ: สหราชอาณาจักร

The measurement of lithium at low concentrations within nuclear power stations has been identified as an application perfectly addressed by a new specialty flame photometer – The BWB Nuclear.

Designed in conjunction with our outstanding distributor in Germany, Mr. Wolfgang Glock, this unique flame photometer has 4 active channels for Lithium detection. Each channel is independently calibrated within the range of interest and this results in an increase of accuracy in keeping with industry demand.

Our simultaneous detection and display from each detection channel satisfies the requirement specifically and the tie of one of the
channels to 2 wire 4-20mA output via supervisory control and data
acquisition (SCADA™) links completes the unit for online use.

Data output via our FP-PC PC application enables data sharing over the internet or intranet for users sharing information.

Clients who utilize our unique “Collection Cup” feature realize online real time analysis. Optional for ‘online’, continuous measurements, the collection cup turns the BWB NUCLEAR FES into an ‘online’, real time, 24/7 analyzer with 4-20mA output linked to a single channel. Via a 2 wire output the BWB FP can seamlessly integrate with any SCADA™ system. Totally unique and available on all models.”


Technical Data

Sample rate 
3.5 – 7ml/min

Tubing materials 
Silicone and Tygon®
 
Required desk space for Instrument  
50cm (H) x 45cm (W) x 45cm (D)*
For Safety reasons the Flame photometer requires 1m of unobstructed space above to allow dissipation of heat from the chimney.

Instrument size 
Nuclear Flame Photometer
51cm (H) x 38cm (W) x 41cm (D)
(20in x 14in x 16in)

Shipping
62cm (H) x 55cm (W) x 47cm (D)
(24in x 22in x 19in)

Weight 
Nuclear Flame Photometer – 15.3kg
Shipping – 25kg

Optimal range
Single point calibration
Li 0.05 – 50ppm 

Multi Point Calibration
Li 0.05 – 1000ppm 

Reproducibility 
<1% Coefficient of variability for 20 consecutive samples over ten minutes at concentrations of 100ppm or less. (After instrument stabilization). 

Chemical Interference 
Na/K/Li = <0.5% to each other when equal in concentration at <100ppm
Limit of detection (LOD) and
limit of quantification (LOQ)  

LOD                    LOQ        
Li – 0.02ppm     Li – 0.05ppm

Time to stability  
Less than 15 seconds after sample
is introduced into the flame

Ions measured  
3 independent Li channels, each providing
up to 14 points of calibration

Interfaces  
USB
0-2.5 volt output (based on sample concentration linked to element of users choice)
Optional 4-20mA output in place of the above
Optional integrated printer
.csv and .pdf generated reports and files via FP-PC software

Recommended minimum warm up
time based on ambient temperature 
 
21ºC – 40 minutes     

Power requirements   
100V – 250V AC at 50 or 60Hz
automatically selecting 

Fuel requirement    
Propane, Butane or Natural Gas*
regulated to 19Bar. Flow rate of 0.4l/min
*with modifications. BWB Technologies recommends either Propane or Butane for optimum results.

Readout  
LCD, four line, alpha numeric, back lit.