With our new “Light Pipe Technology” and multiple calcium detectors systems we are now able to meet the rigid standards of calcium detection in wine as this additive is blended in and then removed out in processing. Of course our potassium and Sodium channels are also specified for this important industry as well.
Used by the more discriminating wine producers in the world, the BWB Wine flame photometer not only offers a superior analytical instrument for the industry but is also a vital tool for optimizing product quality through early detection and measurement of these key elements.
Technical Data
Sample rate
3.5 – 7ml/min
Tubing materials
Silicone and Tygon®
Required desk space for Instrument
50cm (H) x 45cm (W) x 45cm (D)*
For Safety reasons the Flame photometer requires 1m of unobstructed space above to allow dissipation of heat from the chimney.
Instrument size
Wine Flame Photometer
51cm (H) x 38cm (W) x 41cm (D)
(20in x 14in x 16in)
Shipping
62cm (H) x 55cm (W) x 47cm (D)
(24in x 22in x 19in)
Weight
Wine Flame Photometer – 15.3kg
Shipping – 25kg
Optimal range
Single point calibration
Na 0.1 – 60ppm
K 0.05 – 100ppm
Ca 1.0 – 100ppm
Multi Point Calibration
Na 0.1 – 1000ppm
K 0.05 – 1000ppm
Ca 1.0 – 1000ppm
Reproducibility
<1% Coefficient of variability for 20 consecutive samples over ten minutes at concentrations of 100ppm or less. (After instrument stabilisation).
Specificity
Na/K/Li = <0.5% to each other when
equal in concentration at <100ppm
Limit of detection (LOD) and
limit of quantification (LOQ) displayed
LOD LOQ
Na – 0.03ppm Na – 0.1ppm
K – 0.02ppm K – 0.05ppm
Ca – 0.3ppm Ca – 1ppm
Time to stability
Less than 15 seconds after sample
is introduced into the flame
Ions measured
Simultaneous measurement of Na, K, and Ca
Interfaces
USB
0-2.5 volt output (based on sample concentration linked to element of users choice)
Optional 4-20mA output in place of the above
Optional integrated printer
.CSV and .PDF generated reports and files via FP-PC software
Recommended minimum warm up
time based on ambient temperature
21ºC – 40 minutes
Power requirements
100V – 250V AC at 50 or 60Hz
automatically selecting
Fuel requirement
Propane, Butane or Natural Gas*
regulated to 19Bar. Flow rate of 0.4l/min
*with modifications. BWB Technologies recommends either Propane or Butane for optimum results.
Readout
LCD, four line, alpha numeric, back lit.